ACQUIRING AND PROCESSING DATA ON THE EFT-90 SPECTROMETER
| Index |
Acquiring a Spectrum
The PNMR Program
Instrument Parameters
Basic Steps for Data Acquisition
Processing Data
1. Perforing a Fourier Transform
2. Phasing the Spectrum
3. Choosing a region of interest
4. Calibrating the spectrum
| Acuiring a Spectrum |
The NMR computer uses separate programs for acquiring
and processing data. The program used for acquiring data is PNMR; a screenshot
of this program is shown below. If this program is not already loaded,
load it now.
![]() |
Instrument Parameters
Here is a brief explanation of the various parameters:
1. SI represents the size of the dataset; this is the amount of computer memory used to store each file. Typical values are 8192, 16384, or 32768.
2. NS represents the number of scans. Typically 4 or 8 scans are used for an H-1 spectrum; up to several thousand scans may be required to obtain a suitable C-13 spectrum.
3. RG is the receiver gain, or the amount by which the signal obtained from the receiver is amplified. This value ranges from 1 to 100. The more dilute the sample, the greater the receiver gain will have to be in order to obtain a suitable spectrum.
4. RD is the relaxation delay; this is the time the computer waits between the end of data collection and the application of the next pulse. 5. AQ is the acquisition time; this is the length of time the computer collects data after each pulse. This cannot be set directly, but is a function of SI. The larger the size of the dataset, the larger will be the acquisition time. Feel free to experiment with changing SI to see how this influences the acquisition time.
6. N1 represents the nucleus studied; the two most common are H1 and C13.
7. F1 is the transmitter frequency; for an H1 nucleus this has a value of 90.019 MHz. This number will be different if N1 is set to C13 because C13 nuclei come into resonance at a different frequency.
8. W1 represents the spectral width; the default value is 1470.6 Hz. On a 90 MHz spectrometer, 1 ppm = 90 MHz and so this value is equivalent to a spectral width of 1470.6 / 90 = 16.34 ppm. This value can be changed, depending upon the desired spectral width.
9. OF represents the offset frequency; the amount by which the transmitter is moved from the base frequency.
10. PW represents the pulse width; the length of time the radio-frequency pulse is applied. Note that this is given in microseconds; hence the pulse is applied for a very short time.
The lowest portion of the PNMR screen is divided into three windows; the leftmost window is used to enter commands. To change a parameter, type the parameter, followed by a space and the desired value. For example, to change SI from 16386 to 8192 you would type SI 8192 followed by the return key.
Basic Steps for Data Acquisition
To acquire a spectrum using PNMR, you follow the following steps.
1. Change N1 to the nucleus to be studied; type N1 H1 or N1 C13 and press return. If the nucleus is already set to the one you wish to study, you can skip this step.
2. Perform a shim; type shim and press return. The purpose of shimming a sample is to obtain a completely homogenous magnetic filed; shimming adjusts the current in small coils that augment the magnetic field. Shimming varies from sample to sample, depending upon such parameters as the amount of sample in a tube, or small imperfections in the tube itself. Therefore, it is a good idea to perform a shim every time a new sample is inserted in the spectrometer. Poor shimming usually results in broad or asymmetric lines.
3. Set the receiver gain. The shim routine automatically adjusts the receiver gain, and if you are obtaining an H-1 spectrum this value is usually acceptable. If obtaining a C-13 spectrum, you will most likely need to set the gain as high as possible; in this case enter RG 100 and press return.
4. Set the number of scans. If you are obtaining an H-1 spectrum, usually 4 or 8 scans is sufficient. If you are obtaining a C-13 spectrum, you will need more scans, depending upon the concentration of the sample. To set the number of scans to 8, type ns 8 and press return.
5. Type zg and press return. You will be prompted to enter a filename for your dataset. The software is configured to look for data in the data directory; so your filename should be entered in the form data\filename. It is a good idea to get into the habit of using descriptive filenames for your datasets. For example, if you use filenames such as mydata1 and mydata2 you will quickly lose track of your files. Better names would include the sample and type of experiment performed, such as sarah_unknown1_H1 or sarah_unknown1_C13.
In the Fourier transform experiment data is collected
as a function of time rather than frequency. The resulting data is referred
to as a FID or free induction decay. A typical FID is shown below. If the
FID is shown in red the receiver gain is too high; you should re-acquire
the data with a lower receiver gain.
![]() |
You are now ready to process your data.
| Processing Data |
Processing data is performed using the NUTS program. If this program is not already loaded, load it now.
1. Performing the Fourier Transform
The NUTS program has a series of macros for processing
various types of data. A macro is simply a text file that contains a series
of commands; the same effect can be achieved by entering the commands manually,
but is it easier use the macro. From the file menu at the top of the screen,
choose run macro. You will be taken to a dialog box containing a list of
macros. If you have acquired a H-1 spectrum, choose H1.mac; if you have
acquired a C-13 spectrum choose C13.mac. After choosing the proper macro,
you will be taken to a second dialog box asking for the name of the file
to be processed.
![]() |
For processing C-13 spectra, you will be asked to enter a value for the line broadening (LB). This is a function that is applied to the FID prior to processing. Larger values give a better signal-to-noise ratio at the expense of resolution. A good general-purpose value is 1.0. If spectrum contains a number of closely spaced lines; you may wish to enter a smaller value.
At this point you will see a dialog box that lists the
parameters used to acquire the spectrum. If you wish, you may enter a comment
(such as the compound name) or a user name. When finished, click the OK
button on this box.
![]() |
Once you close the "data acauisition parameters"
dialog box shown above, you may see something like the screen below. The
spectrum looks odd because it has not been properly phased. The next step
is to phase the spectrum. The Fourier transform produces an equal number
of absorptive and dispersive data points; the purpose of phasing is to
adjust the way these components are mixed.
![]() |
From the Process menu select Phasing,
and then Phase Correct. If this does not give a satisfactory spectrum,
select Auto Phase from the same menu. This gives the properly
a properly phased spectrum shown below.
![]() |
3. Choosing a Region of Interest
By default, the entire spectral width is shown when a
spectrum is first displayed. However, you may wish to zoom in on a region
of interest. To do this, enter the zoom routine by typing zo (note
that you do not have to press the return key to enter this command); you
should see the cursor change to a crosshair. To select a region of the
spectrum, hold down the left mouse button and drag the cursor over the
region of interest. You will see the spectrum highlighted in red. Then
click the right mouse button; you should see your region of interest expand
to fill the entire screen. If you wish to return to the entire spectrum,
click the right mouse button. When finished, exit the zoom routine by going
to the file menu and choosing exit zo or by simply pressing
the return key.
![]() |
Sometimes the magnetic field of an instrument can vary, and peaks do not always appear in the proper place. For this reason, a chemical shift standard such as TMS (tetramethylsilane) is included in most organic samples. (In fact, the solvent used may already have a small amount of TMS added to it). The TMS should appear as a singlet at a chemical shift of 0.0 ppm. If it occurs in a different position, you will have to adjust the offset so that it does appear at 0.0 ppm.
Suppose, for example, that TMS signal appears at 0.50
ppm instead of 0.00 ppm. To change the offset, go to the view
window and choose spectral parameters. On a 90 MHz spectrometer
1 ppm = 90 Hz, and therefore 0.50 ppm = 45 Hz. Therefore we need to subtract
45 from the offset. If the offset reads 530, for example, enter a value
of 485. This will change the scale on the spectrum so that the TMS signal
has a value of 0.0 ppm.
Alternately, the spectrum can be calibrated using the fo (field
offset) command. To execture this command, return to the PNMR window
and type fo; you will be prompted to first enter the current chemical
shift of the standard, and then then the desired chemical shift. Continuing
from the example above, you would enter 0.50 for the current chemical shift
and 0.0 for the desired chemical shift. After executing this command you
must re-aqcuire and process your data. This method is preferable
to first if large changes must be made to position of the standard.